OEM Manufacturer Faecal Calprotectin Levels - Factory Free sample Rapid test kit / HIV rapid test cassette (whole blood / serum / plasma) MSLRDT007 – Baysen

short description:


Product Detail

Product Tags

Related Video

Feedback (2)

We believe that prolonged expression partnership is really a result of top of the range, value added support, rich encounter and personal contact for Chlamydia Rapid Test Cassette , Hcg Home Rapid Test Strip , Helicobacter Test , Adhering to your small business principle of mutual positive aspects, we have now won superior popularity among our customers because of our best solutions, excellent products and competitive selling prices. We warmly welcome clients from your home and overseas to cooperate with us for common achievement.
OEM Manufacturer Faecal Calprotectin Levels - Factory Free sample Rapid test kit / HIV rapid test cassette (whole blood / serum / plasma) MSLRDT007 – Baysen Detail:

All we do is always involved with our tenet ” Consumer initial, Trust first, devoting within the food stuff packaging and environmental defense for Factory Free sample Rapid test kit / HIV rapid test cassette (whole blood / serum / plasma) MSLRDT007, To improved expand market, we sincerely invite ambitious individuals and providers to hitch as an agent.
All we do is always involved with our tenet ” Consumer initial, Trust first, devoting within the food stuff packaging and environmental defense for , We’ve many years’ experience in hair product production, and our strict QC Team and skilled workers will ensure that we provide you with top hair items with the best hair quality and workmanship. You will get successful business if you choose to cooperate with such a professional manufacturer. Welcome your order cooperation!
Diagnostic Kit for Thyroid Stimulating Hormone immunochromatographic assay)
For in vitro diagnostic use only
Please read this package insert carefully prior to use and strictly follow the instructions. Reliability of assay results cannot be guaranteed if there are any deviations from the instructions in this package insert.

INTENDED USE
Diagnostic Kit for Thyroid Stimulating Hormone (fluorescence immunochromatographic assay) is a fluorescence immunochromatographic assay for the quantitative detection of Thyroid Stimulating Hormone (TSH) in human serum or plasma, which is mainly used in the evaluation of pituitary-thyroid function. All positive sample must be confirmed by other methodologies. This test is intended for healthcare professional use only.

SUMMARY
The main functions of TSH: 1, promote the release of thyroid hormones, 2, promote the synthesis of T4, T3, including strengthening iodine pump activity, enhancing peroxidase activity, promoting the synthesis of thyroid globulin and tyrosine iodide

PRINCIPLE OF THE PROCEDURE
The membrane of the test device is coated with anti TSH antibody on the test region and goat anti rabbit IgG antibody on the control region. Lable pad are coated by fluorescence labeled anti TSH antibody and rabbit IgG in advance. When testing positive sample, the TSH antigen in sample combine with fluorescence labeled anti TSH antibody, and form immune mixture. Under the action of the immunochromatography, the complex flow in the direction of absorbent paper, when complex passed the test region, it combined with anti TSH coating antibody, forms new complex.TSH level is positively correlated with fluorescence signal, and the concentration of TSH in sample can be detected by fluorescence immunoassay assay.

REAGENTS AND MATERIALS SUPPLIED

25T package components
.Test card individually foil pouched with a desiccant 25T
.Sample diluents
.Package insert

MATERIALS REQUIRED BUT NOT PROVIDED
Sample collection container,timer

SAMPLE COLLECTION AND STORAGE
1.The samples tested can be serum, heparin anticoagulant plasma or EDTA anticoagulant plasma.

2.According to standard techniques collect sample. Serum or plasma sample can be kept refrigerated at 2-8℃ for 7days and cryopreservation below -15°C for 6 months.
3.All sample avoid freeze-thaw cycles.

ASSAY PROCEDURE
The test procedure of the instrument see the immunoanalyzer manual. The reagent test procedure is as follows

1.Lay aside all reagents and samples to room temperature.
2.Open the Portable Immune Analyzer(WIZ-A101), enter the account password login according to the operation method of the instrument, and enter the detection interface.
3.Scan the dentification code to confirm the test item.
4.Take out the test card from the foil bag.
5.Insert the test card into the card slot, scan the QR code, and determine the test item.
6.Add 20μL serum or plasma sample to sample diluent, and mix well..
7.Add 80μL sample solution to sample well of the card.
8.Click the “standard test” button, after 15 minutes, the instrument will automatically detect the test card, it can read the results from the display screen of the instrument, and record/print the test results.
9.Refer to the instruction of Portable Immune Analyzer(WIZ-A101).

TEST RESULTS AND INTERPRETATION
.The above data is the result of TSH reagent test, and it is suggested that each laboratory should establish a range of TSH detection values suitable for the population in this region. The above results are for reference only.
.The results of this method are only applicable to the reference ranges established in this method, and there is no direct comparability with other methods.
.Other factors can also cause errors in detection results, including technical reasons, operational errors and other sample factors.

STORAGE AND STABILITY
1.The kit is 18 months shelf-life from the date of manufacture. Store the unused kits at 2-30°C. DO NOT FREEZE. Do not use beyond the expiration date.

2.Do not open the sealed pouch until you are ready to perform a test, and the single-use test is suggested to be used under the required environment (temperature 2-35℃, humidity 40-90%) within 60 mins as quickly as possible.
3.Sample diluent is used immediately after being opened.

WARNINGS AND PRECAUTIONS
.The kit should be sealed and protected against moisture.

.All positive specimens shall be validated by other methodologies.
.All specimens shall be treated as potential pollutant.
.DO NOT use expired reagent.
.DO NOT interchange reagents among kits with different lot No..
.DO NOT reuse test cards and any disposable accessories.
.Misoperation, excessive or little sample can lead to result deviations.

LIMITATION
.As with any assay employing mouse antibodies, the possibility exists for interference by human anti-mouse antibodies (HAMA) in the specimen. Specimens from patients who have received preparations of monoclonal antibodies for diagnosis or therapy may contain HAMA. Such specimens may cause false positive or false negative results.

.This test result is only for clinical reference, should not serve as the only basis for clinical diagnosis and treatment, the patients clinical management should be comprehensive consideration combined with its symptoms, medical history, other laboratory examination, treatment response, epidemiology and other information.
.This reagent is only used for serum and plasma tests. It may not obtain accurate result when used for other samples such as saliva and urine and etc.

PERFORMANCE CHARACTERISTICS

Linearity 0.5μIU/mL to 100μIU/mL relative deviation:-15% to +15%.
    Linear correlation coefficient:(r)≥0.9900
Accuracy The recovery rate shall be within 85% – 115%.
Repeatability CV≤15%
Specificity(None of the substances at the interferent tested interfered in the assay)   

Interferent

Interferent concentration

HCG

2000mIU/mL

FSH

500mIU/mL

LH

500mIU/mL

REFERENCES
1.Hansen JH,et al.HAMA Interference with Murine Monoclonal Antibody-Based Immunoassays[J].J of Clin Immunoassay,1993,16:294-299.
2.Levinson SS.The Nature of Heterophilic Antibodies and the Role in Immunoassay Interference[J].J of Clin Immunoassay,1992,15:108-114.

Key to symbols used:

 t11-1 In Vitro Diagnostic Medical Device
 tt-2 Manufacturer
 tt-71 Store at 2-30℃
 tt-3 Expiration Date
 tt-4 Do Not Reuse
 tt-5 CAUTION
 tt-6 Consult Instructions For Use

Xiamen Wiz Biotech CO.,LTD
Address:3-4 Floor,NO.16 Building,Bio-medical Workshop,2030 Wengjiao West Road,Haicang District,361026,Xiamen,China
Tel:+86-592-6808278
Fax:+86-592-6808279


Product detail pictures:

OEM Manufacturer Faecal Calprotectin Levels - Factory Free sample Rapid test kit / HIV rapid test cassette (whole blood / serum / plasma) MSLRDT007 – Baysen detail pictures

OEM Manufacturer Faecal Calprotectin Levels - Factory Free sample Rapid test kit / HIV rapid test cassette (whole blood / serum / plasma) MSLRDT007 – Baysen detail pictures

OEM Manufacturer Faecal Calprotectin Levels - Factory Free sample Rapid test kit / HIV rapid test cassette (whole blood / serum / plasma) MSLRDT007 – Baysen detail pictures


Related Product Guide:
Do England have what it takes to topple the All Blacks? The Roar | Cpn-Igm
Cardiac Biomarkers Market To Reach $13.3 Billion By 2024: Grand View Research, Inc. | Psa Test Cost

We stick to our enterprise spirit of "Quality, Performance, Innovation and Integrity". We purpose to create a lot more price for our prospects with our rich resources, innovative machinery, experienced workers and great products and services for OEM Manufacturer Faecal Calprotectin Levels - Factory Free sample Rapid test kit / HIV rapid test cassette (whole blood / serum / plasma) MSLRDT007 – Baysen , The product will supply to all over the world, such as: Poland, Tanzania, Mauritius, we're now looking forward to even greater cooperation with overseas customers based on mutual benefits. We are going to work wholeheartedly to improve our products and services. We also promise to work jointly with business partners to elevate our cooperation to a higher level and share success together. Warmly welcome you to visit our factory sincerely.
  • Production management mechanism is completed, quality is guaranteed, high credibility and service let the cooperation is easy, perfect!
    5 Stars By Claire from Guinea - 2016.12.31 14:53
    This enterprise in the industry is strong and competitive, advancing with the times and develop sustainable, we are very pleased to have a opportunity to cooperate!
    5 Stars By Diego from Czech Republic - 2016.10.27 12:12