We now have a highly efficient crew to deal with inquiries from clients. Our intention is "100% shopper pleasure by our merchandise quality, price tag & our staff service" and take pleasure in a very good standing amongst purchasers. With quite a few factories, we can easily provide a wide vary of Psa Test Cost , Follicle Stimulating Rapid Test Kit , Hcv Strip Test , Created solutions with brand price. We attend seriously to produce and behave with integrity, and because of the favor of clients in your own home and overseas in the xxx industry.
New Arrival China 5-part Hematology Analyzer - Diagnostic Kit for Cardiac Troponin I (fluorescence immunochromatographic assay) – Baysen Detail:

Diagnostic Kit for Cardiac Troponin I (fluorescence immunochromatographic assay)
For in vitro diagnostic use only

Please read this package insert carefully prior to use and strictly follow the instructions. Reliability of assay results cannot be guaranteed if there are any deviations from the instructions in this package insert.

INTENDED USE
Diagnostic Kit for Cardiac Troponin I(fluorescence immunochromatographic assay) is a fluorescence immunochromatographic assay for the quantitative detection of Cardiac Troponin I (cTnI) in human serum or plasma, it is used for auxiliary diagnosis of AMI(Acute Myocardial Infarction). All positive sample must be confirmed by other methodologies. This test is intended for healthcare professional use only.

SUMMARY
cTnI levels increased several hours after Myocardial infarction occurred, peaked at 12-16 hours, and remained elevated 4-9 days after Myocardial infarction occurred. The global definition of the third myocardial infarction in 2012:The preferred biomarker-cTn(I or T),has high myocardial tissue specificity and high clinical sensitivity. Changes in the concentration of cTn are essential to the diagnosis of AMI

PRINCIPLE OF THE PROCEDURE
The membrane of the test device is coated with anti cTnI antibody on the test region and goat anti rabbit IgG antibody on the control region. Lable pad are coated by fluorescence labeled anti cTnI antibody and rabbit IgG in advance. When testing positive sample, the cTnI antigen in sample combine with fluorescence labeled anti cTnI antibody, and form immune mixture. Under the action of the immunochromatography, the complex flow in the direction of absorbent paper, when complex passed the test region, it combined with anti cTnI coating antibody, forms new complex.cTnI level is positively correlated with fluorescence signal, and the concentration of cTnI in sample can be detected by fluorescence immunoassay assay.

REAGENTS AND MATERIALS SUPPLIED

25T package components
Test card individually foil pouched with a desiccant 25T
Sample diluents 25T
Package insert 1

MATERIALS REQUIRED BUT NOT PROVIDED
Sample collection container,timer

SAMPLE COLLECTION AND STORAGE
1.The samples tested can be serum, heparin anticoagulant plasma or EDTA anticoagulant plasma.

2.According to standard techniques collect sample. Serum or plasma sample can be kept refrigerated at 2-8℃ for 7days and cryopreservation below -15°C for 6 months.
3.All sample avoid freeze-thaw cycles.

ASSAY PROCEDURE
Please read the instrument operation manual and package insert before testing.

1.Lay aside all reagents and samples to room temperature.
2.Open the Portable Immune Analyzer(WIZ-A101), enter the account password login according to the operation method of the instrument, and enter the detection interface.
3.Scan the dentification code to confirm the test item.
4.Take out the test card from the foil bag.
5.Insert the test card into the card slot, scan the QR code, and determine the test item.
6.Add 40μL serum or plasma sample to sample diluent, and mix well..
7.Add 80μL sample solution to sample well of the card.
8.Click the “standard test” button, after 15 minutes, the instrument will automatically detect the test card, it can read the results from the display screen of the instrument, and record/print the test results.
9.Refer to the instruction of Portable Immune Analyzer(WIZ-A101).

EXPECTED VALUES
cTnI <0.3ng/mL

It is recommended that each laboratory establish its own normal range representing its patient population.

TEST RESULTS AND INTERPRETATION
.The above data is the result of cTnI reagent test, and it is suggested that each laboratory should establish a range of cTnI detection values suitable for the population in this region. The above results are for reference only.

.The results of this method are only applicable to the reference ranges established in this method, and there is no direct comparability with other methods.
.Other factors can also cause errors in detection results, including technical reasons, operational errors and other sample factors.

STORAGE AND STABILITY
1.The kit is 18 months shelf-life from the date of manufacture. Store the unused kits at 2-30°C. DO NOT FREEZE. Do not use beyond the expiration date.

2.Do not open the sealed pouch until you are ready to perform a test, and the single-use test is suggested to be used under the required environment (temperature 2-35℃, humidity 40-90%) within 60 mins as quickly as possible.
3.Sample diluent is used immediately after being opened.

WARNINGS AND PRECAUTIONS
.The kit should be sealed and protected against moisture.

.All positive specimens shall be validated by other methodologies.
.All specimens shall be treated as potential pollutant.
.DO NOT use expired reagent.
.DO NOT interchange reagents among kits with different lot No..
.DO NOT reuse test cards and any disposable accessories.
.Misoperation, excessive or little sample can lead to result deviations.

LIMITATION
.As with any assay employing mouse antibodies, the possibility exists for interference by human anti-mouse antibodies (HAMA) in the specimen. Specimens from patients who have received preparations of monoclonal antibodies for diagnosis or therapy may contain HAMA. Such specimens may cause false positive or false negative results.

.This test result is only for clinical reference, should not serve as the only basis for clinical diagnosis and treatment, the patients clinical management should be comprehensive consideration combined with its symptoms, medical history, other laboratory examination, treatment response, epidemiology and other information.
.This reagent is only used for serum and plasma tests. It may not obtain accurate result when used for other samples such as saliva and urine and etc.

PERFORMANCE CHARACTERISTICS

Linearity 0.1ng/mL to 40ng/mL relative deviation:-15% to +15%.
Linear correlation coefficient:(r)≥0.9900
Accuracy The recovery rate shall be within 85% – 115%.
Repeatability CV≤15%
Specificity(None of the substances at the interferent tested interfered in the assay)

Interferent

Interferent concentration

sTnI

1000μg/L

cTnT

1000μg/L

ABP

1000μg/L

CK-MB

1000μg/L

cTnC

1000μg/L

sTnT

1000μg/L

MYO

1000μg/L

REFERENCES

1.Hansen JH,et al.HAMA Interference with Murine Monoclonal Antibody-Based Immunoassays[J].J of Clin Immunoassay,1993,16:294-299.
2.Levinson SS.The Nature of Heterophilic Antibodies and the Role in Immunoassay Interference[J].J of Clin Immunoassay,1992,15:108-114.

Key to symbols used:

 t11-1 In Vitro Diagnostic Medical Device
 tt-2 Manufacturer
 tt-71 Store at 2-30℃
 tt-3 Expiration Date
 tt-4 Do Not Reuse
 tt-5 CAUTION
 tt-6 Consult Instructions For Use

Xiamen Wiz Biotech CO.,LTD
Address:3-4 Floor,NO.16 Building,Bio-medical Workshop,2030 Wengjiao West Road,Haicang District,361026,Xiamen,China
Tel:+86-592-6808278
Fax:+86-592-6808279


Product detail pictures:

New Arrival China 5-part Hematology Analyzer - Diagnostic Kit for Cardiac Troponin I (fluorescence immunochromatographic assay) – Baysen detail pictures


Related Product Guide:
It’s time for a new Europe-wide strategy on prostate cancer | P24 Test Strips
Swell High-Protein Ice Cream at Trader Joe’s | Cpn-Igm

We usually think and practice corresponding for the change of circumstance, and grow up. We aim at the achievement of a richer mind and body as well as the living for New Arrival China 5-part Hematology Analyzer - Diagnostic Kit for Cardiac Troponin I (fluorescence immunochromatographic assay) – Baysen , The product will supply to all over the world, such as: Dubai, Cologne, Monaco, Insisting over the high-quality generation line management and prospects guide provider, we've made our resolution to offer our shoppers using the initially stage purchasing and soon after provider working experience. Preserving the prevailing helpful relations with our prospects, we even now innovate our product lists the many time to meet up with the brand new wants and stick to the latest trend of this business in Ahmedabad. We're ready to facial area the difficulties and make the transform to grasp many of the possibilities in international trade.
  • This company has a lot of ready-made options to choose and also could custom new program according to our demand, which is very nice to meet our needs.
    5 Stars By Dominic from Birmingham - 2016.12.09 14:01
    We are really happy to find such a manufacturer that ensuring product quality at the same time the price is very cheap.
    5 Stars By Lorraine from luzern - 2015.08.12 12:27