We thinks what clients think, the urgency of urgency to act from the interests of a client position of theory, allowing for greater quality, lessen processing costs, price ranges are much more reasonable, won the new and outdated shoppers the support and affirmation for Diagnostic Kit For Total Thyroxine , Diagnostic Kit For Pro-Brain Natriuretic Peptide , Diagnostic Kit For Stimulating Hormone , Our goal is to create Win-win situation with our customers. We believe we will be your best choice. "Reputation First, Customers Foremost. "Waiting for your inquiry.
Factory Promotional Poct Immunoassay Analyzer Ce - Diagnostic Kit for D-Dimer (fluorescence immunochromatographic assay) – Baysen Detail:

Diagnostic Kit for D-Dimer(fluorescence immunochromatographic assay)
For in vitro diagnostic use only

Please read this package insert carefully prior to use and strictly follow the instructions. Reliability of assay results cannot be guaranteed if there are any deviations from the instructions in this package insert.

INTENDED USE
Diagnostic Kit for D-Dimer(fluorescence immunochromatographic assay) is a fluorescence immunochromatographic assay for the quantitative detection of D-Dimer (D-D) in human plasma, it is used for the diagnosis of venous thrombosis, disseminated intravascular coagulation, and the monitoring of thrombolytic therapy .All positive sample must be confirmed by other methodologies. This test is intended for healthcare professional use only.

SUMMARY
D-D reflects fibrinolytic function.The reasons for the increase of D-D:1.Secondary hyperfibrinolysis, such as hypercoagulation, disseminated intravascular coagulation, renal disease, organ transplant rejection, thrombolytic therapy, etc. 2.There are activated thrombus formation and fibrinolysis activities in vessels; 3.Myocardial infarction, cerebral infarction, pulmonary embolism, venous thrombosis, surgery, tumor,diffuse intravascular coagulation, infection and tissue necrosis, etc

PRINCIPLE OF THE PROCEDURE
The membrane of the test device is coated with anti D-D antibody on the test region and goat anti rabbit IgG antibody on the control region. Lable pad are coated by fluorescence labeled anti D-D antibody and rabbit IgG in advance. When testing positive sample, the D-D antigen in sample combine with fluorescence labeled anti D-D antibody, and form immune mixture. Under the action of the immunochromatography, the complex flow in the direction of absorbent paper, when complex passed the test region, it combined with anti D-D coating antibody, forms new complex. D-D level is positively correlated with fluorescence signal, and the concentration of D-D in sample can be detected by fluorescence immunoassay assay.

REAGENTS AND MATERIALS SUPPLIED

25T package components
Test card individually foil pouched with a desiccant 25T
Sample diluents 25T
Package insert 1

MATERIALS REQUIRED BUT NOT PROVIDED
Sample collection container,timer

SAMPLE COLLECTION AND STORAGE
.The samples tested can be heparin anticoagulant plasma or EDTA anticoagulant plasma.

.According to standard techniques collect sample. Serum or plasma sample can be kept refrigerated at 2-8℃ for 7days and cryopreservation below -15°C for 6 months.
.All sample avoid freeze-thaw cycles.

ASSAY PROCEDURE
Please read the instrument operation manual and package insert before testing.

1.Lay aside all reagents and samples to room temperature.
2.Open the Portable Immune Analyzer(WIZ-A101), enter the account password login according to the operation method of the instrument, and enter the detection interface.
3.Scan the dentification code to confirm the test item.
4.Take out the test card from the foil bag.
5.Insert the test card into the card slot, scan the QR code, and determine the test item.
6.Add 40μL serum or plasma sample to sample diluent, and mix well..
7.Add 80μL sample solution to sample well of the card.
8.Click the “standard test” button, after 15 minutes, the instrument will automatically detect the test card, it can read the results from the display screen of the instrument, and record/print the test results.
9.Refer to the instruction of Portable Immune Analyzer(WIZ-A101).

EXPECTED VALUES
D-D <0.5mg/L

It is recommended that each laboratory establish its own normal range representing its patient population.

TEST RESULTS AND INTERPRETATION
.The above data is the result of D-D reagent test, and it is suggested that each laboratory should establish a range of D-D detection values suitable for the population in this region. The above results are for reference only.

.The results of this method are only applicable to the reference ranges established in this method, and there is no direct comparability with other methods.
.Other factors can also cause errors in detection results, including technical reasons, operational errors and other sample factors.

STORAGE AND STABILITY
1.The kit is 18 months shelf-life from the date of manufacture. Store the unused kits at 2-30°C. DO NOT FREEZE. Do not use beyond the expiration date.

2.Do not open the sealed pouch until you are ready to perform a test, and the single-use test is suggested to be used under the required environment (temperature 2-35℃, humidity 40-90%) within 60 mins as quickly as possible.
3.Sample diluent is used immediately after being opened.

WARNINGS AND PRECAUTIONS
1.The kit should be sealed and protected against moisture.

2.All positive specimens shall be validated by other methodologies.
3.All specimens shall be treated as potential pollutant.
4.DO NOT use expired reagent.
5.DO NOT interchange reagents among kits with different lot No..
6.DO NOT reuse test cards and any disposable accessories.
7.Misoperation, excessive or little sample can lead to result deviations.

LIMITATION
.As with any assay employing mouse antibodies, the possibility exists for interference by human anti-mouse antibodies (HAMA) in the specimen. Specimens from patients who have received preparations of monoclonal antibodies for diagnosis or therapy may contain HAMA. Such specimens may cause false positive or false negative results.

.This test result is only for clinical reference, should not serve as the only basis for clinical diagnosis and treatment, the patients clinical management should be comprehensive consideration combined with its symptoms, medical history, other laboratory examination, treatment response, epidemiology and other information.
.This reagent is only used for serum and plasma tests. It may not obtain accurate result when used for other samples such as saliva and urine and etc.

PERFORMANCE CHARACTERISTICS

Linearity 0.2mg/L to10mg/L relative deviation:-15% to +15%.
Linear correlation coefficient:(r)≥0.9900
Accuracy The recovery rate shall be within 85% – 115%.
Repeatability CV≤15%
Specificity(None of the substances at the interferent tested interfered in the assay)

Interferent

Interferent concentration

FDP

120mg/L

VC

2000mg/L

Barbituric acid

100mg/L

REFERENCES
1.Hansen JH,et al.HAMA Interference with Murine Monoclonal Antibody-Based Immunoassays[J].J of Clin Immunoassay,1993,16:294-299.

2.Levinson SS.The Nature of Heterophilic Antibodies and the Role in Immunoassay Interference[J].J of Clin Immunoassay,1992,15:108-114.

Key to symbols used:

 t11-1 In Vitro Diagnostic Medical Device
 tt-2 Manufacturer
 tt-71 Store at 2-30℃
 tt-3 Expiration Date
 tt-4 Do Not Reuse
 tt-5 CAUTION
 tt-6 Consult Instructions For Use

Xiamen Wiz Biotech CO.,LTD
Address:3-4 Floor,NO.16 Building,Bio-medical Workshop,2030 Wengjiao West Road,Haicang District,361026,Xiamen,China
Tel:+86-592-6808278
Fax:+86-592-6808279


Product detail pictures:

Factory Promotional Poct Immunoassay Analyzer Ce - Diagnostic Kit for D-Dimer (fluorescence immunochromatographic assay) – Baysen detail pictures


Related Product Guide:
Global Enzyme Linked Immunosorbent Assay (ELISA) Market to Surpass US$ 2.5 Billion by 2025 | Psa Test Cost
‘My life was probably saved by a £40 blood test for prostate cancer’ | Calprotectin Elisa Kit

Adhering into the basic principle of "quality, assistance, effectiveness and growth", we have attained trusts and praises from domestic and worldwide client for Factory Promotional Poct Immunoassay Analyzer Ce - Diagnostic Kit for D-Dimer (fluorescence immunochromatographic assay) – Baysen , The product will supply to all over the world, such as: Curacao, Guinea, Sri Lanka, We maintain long-term efforts and self-criticism, which helps us and improvement constantly. We strive to improve customer efficiency to save costs for customers. We do our best to improve the quality of product. We will not live up to the historic opportunity of the times.
  • This is a honest and trustworthy company, technology and equipment are very advanced and the prodduct is very adequate, there is no worry in the suppliment.
    5 Stars By Grace from Guyana - 2016.11.12 12:31
    The after-sale warranty service is timely and thoughtful, encounter problems can be resolved very quickly, we feel reliable and secure.
    5 Stars By Nicole from Bangladesh - 2015.03.03 13:09