Our items are commonly identified and trusted by people and can fulfill repeatedly altering economic and social wants of Crp Blood Test Kit , Hcg Pregnancy Test Kit , Calprotectin Stool Test Normal Range , How about to start your good organization with our corporation? We are all set, properly trained and fulfilled with pride. Let’s start our new business enterprise with new wave.
China New Product Malaria Pan+pf Test Kit - Diagnostic Kit for Total Triiodothyronine (fluorescence immunochromatographic assay) – Baysen Detail:

Diagnostic Kit for Total Triiodothyronine(fluorescence immunochromatographic assay)
For in vitro diagnostic use only

Please read this package insert carefully prior to use and strictly follow the instructions. Reliability of assay results cannot be guaranteed if there are any deviations from the instructions in this package insert.

INTENDED USE
Diagnostic Kit for Total Triiodothyronine (fluorescence immunochromatographic assay) is a fluorescence immunochromatographic assay for the quantitative detection of Total Triiodothyronine (TT3) in human serum or plasma, which is mainly used to evaluate thyroid function.It is an auxiliary diagnosis reagent.All positive sample must be confirmed by other methodologies. This test is intended for healthcare professional use only.

SUMMARY
Triiodothyronine(T3) molecular weight 651D. It is the main active form of thyroid hormone. Total T3(Total T3, TT3) in the serum is divided into binding and free types. 99.5 % of TT3 binds to serum Thyroxine Binding Proteins(TBP), and free T3(Free T3) accounts for 0.2 to 0.4 %. T4 and T3 participate in maintaining and regulating the body’s metabolic function.TT3 measurements are used to evaluate thyroid functional status and diagnosis of diseases. Clinical TT3 is a reliable indicator for the diagnosis and efficacy observation of hyperthyroidism and hypothyroidism.The determination of T3 is more significant for the diagnosis of hyperthyroidism than T4.

PRINCIPLE OF THE PROCEDURE

The membrane of the test device is coated with the conjugate of BSA and T3 on the test region and goat anti rabbit IgG antibody on the control region. Marker pad are coated by fluorescence mark anti T3 antibody and rabbit IgG in advance. When testing sample, TT3 in sample combine with fluorescence marked anti T3 antibody, and form immune mixture. Under the action of the immunochromatography, the complex flow in the direction of absorbent paper, when complex passed the test region, The free fluorescent marker will be combined with T3 on the membrane.The concentration of TT3 is negative correlation for fluorescence signal, and the concentration of TT3 in sample can be detected by fluorescence immunoassay assay.

REAGENTS AND MATERIALS SUPPLIED

25T package components
.Test card individually foil pouched with a desiccant 25T
.A solution 25T
.B solution 1
.Package insert 1

MATERIALS REQUIRED BUT NOT PROVIDED
Sample collection container,timer

SAMPLE COLLECTION AND STORAGE
1.The samples tested can be serum, heparin anticoagulant plasma or EDTA anticoagulant plasma.

2.According to standard techniques collect sample. Serum or plasma sample can be kept refrigerated at 2-8℃ for 7days and cryopreservation below -15°C for 6 months.
3.All sample avoid freeze-thaw cycles.

ASSAY PROCEDURE
The test procedure of the instrument see the immunoanalyzer manual. The reagent test procedure is as follows

1.Lay aside all reagents and samples to room temperature.
2.Open the Portable Immune Analyzer(WIZ-A101), enter the account password login according to the operation method of the instrument, and enter the detection interface.
3.Scan the dentification code to confirm the test item.
4.Take out the test card from the foil bag.
5.Insert the test card into the card slot, scan the QR code, and determine the test item.
6.Add 30μL serum or plasma sample to A solution, and mix well.
7.Add 20μL B solution to the above mixture, and mix well.
8.Leave the mixture for 20 minutes.
9.Add 80μL mixture to sample well of the card.
10.Click the “standard test” button,after 10 minutes, the instrument will automatically detect the test card, it can read the results from the display screen of the instrument, and record/print the test results.
11.Refer to the instruction of Portable Immune Analyzer(WIZ-A101).

EXPECTED VALUES

TT3 normal range:0.5-2.5ng/mL
It is recommended that each laboratory establish its own normal range representing its patient population.

TEST RESULTS AND INTERPRETATION
.The above data is the reference interval established for the detection data of this kit, and it is suggested that each laboratory should establish a reference interval for the relevant clinical significance of the population in this region.

.The concentration of TT3 is higher than the reference range, and the physiological changes or stress response should be excluded.Indeed abnormal, should combine clinical symptom diagnosis.
.The results of this method are only applicable to the reference range established by this method, and the results are not directly comparable with other methods.
.Other factors can also cause errors in detection results, including technical reasons, operational errors and other sample factors.

STORAGE AND STABILITY
1.The kit is 18 months shelf-life from the date of manufacture. Store the unused kits at 2-30°C. DO NOT FREEZE. Do not use beyond the expiration date.

2.Do not open the sealed pouch until you are ready to perform a test, and the single-use test is suggested to be used under the required environment (temperature 2-35℃, humidity 40-90%) within 60 mins as quickly as possible.
3.Sample diluent is used immediately after being opened.

WARNINGS AND PRECAUTIONS
.The kit should be sealed and protected against moisture.

.All positive specimens shall be validated by other methodologies.
.All specimens shall be treated as potential pollutant.
.DO NOT use expired reagent.
.DO NOT interchange reagents among kits with different lot no..
.DO NOT reuse test cards and any disposable accessories.
.Misoperation, excessive or little sample can lead to result deviations.

LIMITATION
.As with any assay employing mouse antibodies, the possibility exists for interference by human anti-mouse antibodies (HAMA) in the specimen. Specimens from patients who have received preparations of monoclonal antibodies for diagnosis or therapy may contain HAMA. Such specimens may cause false positive or false negative results.

.This test result is only for clinical reference, should not serve as the only basis for clinical diagnosis and treatment, the patients clinical management should be comprehensive consideration combined with its symptoms, medical history, other laboratory examination, treatment response, epidemiology and other information.
.This reagent is only used for serum and plasma tests. It may not obtain accurate result when used for other samples such as saliva and urine and etc.

PERFORMANCE CHARACTERISTICS

Linearity 0.25 ng/mL to 10 ng/mL relative deviation:-15% to +15%.
Linear correlation coefficient:(r)≥0.9900
Accuracy The recovery rate shall be within 85% – 115%.
Repeatability CV≤15%
Specificity(None of the substances at the interferent tested interfered in the assay) Interferent Interferent concentration
Hemoglobin 200μg/mL
transferrin 100μg/mL
Horseradish Peroxidase 2000μg/mL
rT3 100ng/mL
T4 200ng/mL

REFERENCES
1.Hansen JH,et al.HAMA Interference with Murine Monoclonal Antibody-Based Immunoassays[J].J of Clin Immunoassay,1993,16:294-299.

2.Levinson SS.The Nature of Heterophilic Antibodies and the Role in Immunoassay Interference[J].J of Clin Immunoassay,1992,15:108-114.

Key to symbols used:

 t11-1 In Vitro Diagnostic Medical Device
 tt-2 Manufacturer
 tt-71 Store at 2-30℃
 tt-3 Expiration Date
 tt-4 Do Not Reuse
 tt-5 CAUTION
 tt-6 Consult Instructions For Use

Xiamen Wiz Biotech CO.,LTD
Address:3-4 Floor,NO.16 Building,Bio-medical Workshop,2030 Wengjiao West Road,Haicang District,361026,Xiamen,China
Tel:+86-592-6808278
Fax:+86-592-6808279


Product detail pictures:

China New Product Malaria Pan+pf Test Kit - Diagnostic Kit for Total Triiodothyronine  (fluorescence immunochromatographic assay) – Baysen detail pictures


Related Product Guide:
Silencing MicroRNA-155 Attenuates Cardiac Injury and Dysfunction in Viral Myocarditis via Promotion of M2 Phenotype Polarization of Macrophages | Cpn-Igm
TXNIP/TRX/NF-κB and MAPK/NF-κB pathways involved in the cardiotoxicity induced by Venenum Bufonis in rats | Diagnostic Kit For Isoenzyme Mb Of C Reatine Kinase

"Control the standard by the details, show the power by quality". Our organization has strived to establish a highly efficient and stable employees team and explored an effective high-quality command method for China New Product Malaria Pan+pf Test Kit - Diagnostic Kit for Total Triiodothyronine (fluorescence immunochromatographic assay) – Baysen , The product will supply to all over the world, such as: Nepal, Armenia, Bangalore, Please feel cost-free to send us your specifications and we'll respond to you asap. We've got a professional engineering team to serve for the every single detailed needs. Free samples may be sent for you personally to know far more facts. So that you can meet your desires, please really feel cost-free to contact us. You could send us emails and call us straight. Additionally, we welcome visits to our factory from all over the world for much better recognizing of our corporation. nd merchandise. In our trade with merchants of several countries, we often adhere to the principle of equality and mutual advantage. It is our hope to market, by joint efforts, both trade and friendship to our mutual benefit. We look forward to getting your inquiries.
  • The factory technical staff gave us a lot of good advice in the cooperation process, this is very good, we are very grateful.
    5 Stars By Margaret from Uruguay - 2015.11.11 19:52
    We feel easy to cooperate with this company, the supplier is very responsible, thanks.There will be more in-depth cooperation.
    5 Stars By Renee from Kuwait - 2016.05.31 13:26
    Write your message here and send it to us